CPAP Vygon Boussignac range
CPAP Vygon Boussignac
Popis
Respiratory assistance device for patients with spontaneous ventilation allowing maintenance in the airways of a pressure greater than atmospheric… Přečtěte si více
Indikace
- Acute pulmonary oedema (APO)
- Dyspnea and acute respiratory distress
- Severe asthma
- Fibroscopic procedures
- Oxygenation (post-operative) in patients at risk of hypoxia (such as cardiothoracic surgery and abdominal surgery in obese patients)
- Post-extubation weaning from mechanical ventilation
Popis
Respiratory assistance device for patients with spontaneous ventilation allowing maintenance in the airways of a pressure greater than atmospheric pressure during the entire respiratory cycle.The CPAP mode of ventilation allows:
- To avoid intubation and mechanical ventilation and associated complications (prolonged hospitalization, infections, etc.)
- To allow a safe extubation postoperatively for patients at risk of hypoxemia and thus avoid mechanical ventilation and associated complications
- To contribute to the benefit of weaning from mechanical ventilation and avoid associated complications
The Boussignac CPAP and its manometer connector is a device for connection to a face mask with its 22 mm male connector, the crenellated upper part remaining open to the air.
This device comprises 2 lateral ports:
- the upper port is preconnected to an extension tube (length 200 cm) for connection to an air or oxygen source
- the bottom port (colourless translucent) allows controlling pressure (use manometer code 527.01).
How it works:
The gases administered to the patient (air, air-oxygen mixture or pure oxygen) pass through the extension tube and arrive in a circular chamber where the gases can only escape into the central zone passing through 4 micro-channels. At that moment, the injection speed of the gases is greatly increased. Deflectors at the exit site of the micro-channels eject the gases arriving at high speed towards the opposite wall.
The collision of gas molecules between themselves generates a turbulence which creates a virtual valve and simultaneously the speed of gases is transformed in pressure. This pressure depends directly on the flowrate: if the flowrate (the amount of gas injected) is increased, the pressure will increase; if the flowrate is decreased, the pressure will decrease.
N.B. It is possible to connect the Boussignac CPAP to an endotracheal tube or to a tracheostomy tube using the adaptor code 555.01 together with an artificial nose such as Hygroflux code 501.01.
Reference a funkce
Balení | |||
---|---|---|---|
Favourites | Kód | Jednotky/balení | Jednotky/karton |
5570.13 | 5 | 60 |
Další informace
-
Obsahuje latex Ne
-
Obsahuje živočišný produkt Ne
-
Bez pyrogenu Ne
Dokumentace
-
Airway management catalogue
-
Brochure CPAP Product
-
Návod k použití